Touch-less Heartbeat Detection and Measurement-based Cardiopulmonary Modeling
Résumé
This paper presents a system for touch-less heartbeat detection and a cardiopulmonary signal modeling approach. Using a vector network analyzer, a microwave system is tested in detecting the heartbeat signal at a distance of 1-m from a person. The proposed system shows the ability of detecting the heartbeat signals with the possibility of tuning both frequency and power; measurements are performed at 2.4, 5.8, 10, 16, and 60 GHz, as well as at different power levels between 0 and -27 dBm. Based on real measurements performed for both respiration and heart beatings, a model representing the cardiopulmonary activity is presented. The heartbeat rate and the heart rate variability are extracted from the modeling signal using classic and wavelet filters, for SNR between 0 and -20 dB.