First inappropriate implantable cardioverter defibrillator therapy is often due to inaccurate device programming: analysis of the French OPERA registry. - Université de Rennes Accéder directement au contenu
Article Dans Une Revue EP-Europace Année : 2012

First inappropriate implantable cardioverter defibrillator therapy is often due to inaccurate device programming: analysis of the French OPERA registry.

Antoine Leenhardt
  • Fonction : Auteur correspondant
  • PersonId : 948448

Connectez-vous pour contacter l'auteur
Pascal Defaye
Elisabeth Mouton
  • Fonction : Auteur
Marc Delay
  • Fonction : Auteur
Nicolas Delarche
  • Fonction : Auteur
Jean-Marc Dupuis
  • Fonction : Auteur
Olivier Bizeau
  • Fonction : Auteur
Saida Cheggour
  • Fonction : Auteur
Dominique Babuty
  • Fonction : Auteur

Résumé

AIMS: Inappropriate therapy delivered by implantable cardioverter defibrillators (ICDs) remains a challenge. The OPERA registry measured the times to, and studied the determinants of, first appropriate (FAT) and inappropriate (FIT) therapies delivered by single-, dual- and triple-chamber [cardiac resynchronization therapy defibrillator (CRT-D)] ICD. METHODS AND RESULTS: We entered 636 patients (mean age = 62.0 ± 13.5 years; 88% men) in the registry, of whom 251 received single-, 238 dual-, and 147 triple-chamber ICD, for primary (30.5%) or secondary (69.5%) indications. We measured times to FAT and FIT as a function of multiple clinical characteristics, examined the effects of various algorithm components on the likelihood of FAT and FIT delivery, and searched for predictors of FAT and FIT. Over 22.8 ± 8.8 months of observation, 184 patients (28.9%) received FAT and 70 (11.0%) received FIT. Ventricular tachycardia (VT) was the trigger of 88% of FAT, and supraventricular tachycardia was the trigger of 91% of FIT. The median times to FIT (90 days; range 49-258) and FAT (171 days; 50-363) were similar. The rate of FAT was higher (P <0.001) in patients treated for secondary than primary indications, while that of FIT were similar in both groups. Out of 57 analysable FIT, 27 (47.4%) could have been prevented by fine tuning the device programming like the sustained rate duration or the VT discrimination algorithm. CONCLUSIONS: First inappropriate therapy occurred in 11% of 636 ICD recipients followed for ∼2 years. Nearly 50% of FIT could have been prevented by improving device programming.

Dates et versions

hal-00904932 , version 1 (15-11-2013)

Identifiants

Citer

Antoine Leenhardt, Pascal Defaye, Elisabeth Mouton, Marc Delay, Nicolas Delarche, et al.. First inappropriate implantable cardioverter defibrillator therapy is often due to inaccurate device programming: analysis of the French OPERA registry.. EP-Europace, 2012, 14 (10), pp.1465-74. ⟨10.1093/europace/eus144⟩. ⟨hal-00904932⟩
108 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More